Παρακαλώ χρησιμοποιήστε αυτό το αναγνωριστικό για να παραπέμψετε ή να δημιουργήσετε σύνδεσμο προς αυτό το τεκμήριο: https://hdl.handle.net/20.500.14279/32603
Τίτλος: A Predictive Toxicokinetic Model for Nickel Leaching from Vascular Stents
Συγγραφείς: Giakoumi, Matheos 
Stephanou, Pavlos S. 
Kokkinidou, Despoina 
Papastefanou, Chara 
Anayiotos, Andreas 
Kapnisis, Konstantinos 
Major Field of Science: Engineering and Technology
Field Category: Mechanical Engineering
Λέξεις-κλειδιά: physiologically based toxicokinetic (PBTK) models;mouse stent implantation model;multiobjective optimization;nickel leaching;vascular stents;toxicological risk assessment
Ημερομηνία Έκδοσης: 8-Απρ-2024
Πηγή: ACS biomaterials science and engineering, 2024, vol. 10, iss. 4, pp. 2534–2551
Volume: 10
Issue: 4
Start page: 2534
End page: 2551
Περιοδικό: ACS Biomaterials Science and Engineering 
Περίληψη: In vitro testing methods offer valuable insights into the corrosion vulnerability of metal implants and enable prompt comparison between devices. However, they fall short in predicting the extent of leaching and the biodistribution of implant byproducts under in vivo conditions. Physiologically based toxicokinetic (PBTK) models are capable of quantitatively establishing such correlations and therefore provide a powerful tool in advancing nonclinical methods to test medical implants and assess patient exposure to implant debris. In this study, we present a multicompartment PBTK model and a simulation engine for toxicological risk assessment of vascular stents. The mathematical model consists of a detailed set of constitutive equations that describe the transfer of nickel ions from the device to peri-implant tissue and circulation and the nickel mass exchange between blood and the various tissues/organs and excreta. Model parameterization was performed using (1) in-house-produced data from immersion testing to compute the device-specific diffusion parameters and (2) full-scale animal in situ implantation studies to extract the mammalian-specific biokinetic functions that characterize the time-dependent biodistribution of the released ions. The PBTK model was put to the test using a simulation engine to estimate the concentration-time profiles, along with confidence intervals through probabilistic Monte Carlo, of nickel ions leaching from the implanted devices and determine if permissible exposure limits are exceeded. The model-derived output demonstrated prognostic conformity with reported experimental data, indicating that it may provide the basis for the broader use of modeling and simulation tools to guide the optimal design of implantable devices in compliance with exposure limits and other regulatory requirements.
URI: https://hdl.handle.net/20.500.14279/32603
ISSN: 23739878
DOI: 10.1021/acsbiomaterials.3c01436
Rights: Attribution-NonCommercial-NoDerivatives 4.0 International
Type: Article
Affiliation: Cyprus University of Technology 
Cp Foodlab Ltd. 
Funding: This work was cofunded by the European Regional Development Fund and the Republic of Cyprus through the Research and Innovation Foundation (Project: POST-DOC/0916/0237).
Publication Type: Peer Reviewed
Εμφανίζεται στις συλλογές:Άρθρα/Articles

Αρχεία σε αυτό το τεκμήριο:
Αρχείο Περιγραφή ΜέγεθοςΜορφότυπος
giakoumi-et-al-2024-a-predictive-toxicokinetic-model-for-nickel-leaching-from-vascular-stents.pdfOpen Access9.88 MBAdobe PDFΔείτε/ Ανοίξτε
CORE Recommender
Δείξε την πλήρη περιγραφή του τεκμηρίου

Page view(s)

57
Last Week
0
Last month
5
checked on 4 Οκτ 2024

Download(s) 50

63
checked on 4 Οκτ 2024

Google ScholarTM

Check

Altmetric


Αυτό το τεκμήριο προστατεύεται από άδεια Άδεια Creative Commons Creative Commons