  4. Evaluation of a flat, rectangular MR compatible ultrasonic transducer for cardiac applications
  • Details

Evaluation of a flat, rectangular MR compatible ultrasonic transducer for cardiac applications

Date Issued
May 2023
Author(s)
Sotiriou, Michalis  
Advisor
Damianou, Christakis A.  
Abstract
High Intensity Focused Ultrasound (HIFU) is a non-invasive therapeutic technique that can cause localized hyperthermia at predictable depths without injuring intervening tissues. Atherosclerosis is a condition that develops when a substance called plaque builds up in the walls of the arteries and is related with numerous problems on heart and blood vessels. HIFU can be used to ablate atherosclerotic plaque from these arteries and avoid fatal problems such as heart failure and stroke.
In this study the main objective is the design of an in vitro model to assess the thermal ablation of atherosclerotic plaque using different parameters of intravascular transducers. The experiments were initially conducted in different phantoms and in freshly excised turkey tissue phantom. A flat rectangular transducer was assessed in an arterial atherosclerotic plaque phantom which was created in the laboratory with a very low cost. Τhe proposed phantom mimics the human atherosclerotic plaque. Atherosclerotic plaque is mainly consists of fibrous tissue, lipid core, calcium and necrotic core. Different percentages of those components would categorize different types of plaques. Agar was used for mimicking fibrous tissue, gypsum for calcium and butter for lipid core. The only limitation is that macrophages were not used for mimicking the necrotic core. The artery phantom follows the elastic properties of an artery. The amount of plaque removal was evaluated visually and using an X-Ray system. The main complications of thermal ablation were the increased artery temperature above a safe level (1 ℃) and the rupture of atherosclerotic plaque during the ablation. Furthermore, different parameters of intravascular transducers were assessed with different composition percentages of atherosclerotic plaque. The frequency was 4.0 MHz and acoustical power and sonication time were between 6–15 W and 15–30 s respectively for achieving the optimum combination. Moreover, the specific flat, rectangular transducer is MR compatible and thermometry images were obtained during this evaluation. The intended application is to use it for atherosclerotic plaque ablation using a single element transducer. Finally, an evaluation of acoustic and thermal properties of plaque phantoms to test their suitability mainly for ultrasound imaging and therapy was performed. This is very important for the effective implementation of ultrasound not only in diagnosis, but especially for therapy. The evaluation included measurements of the acoustic propagation speed using the pulse-echo technique, the ultrasonic attenuation coefficient using through transmission immersion technique, and the absorption coefficient. Moreover, thermal properties (thermal conductivity, volumetric specific heat capacity and thermal diffusivity) were measured with the transient method using a needle probe. The mean value of acoustic and thermal properties and their standard deviation of plaque phantoms were 1523±23 m/s for acoustic speed, 0.50±0.02 W/mK for thermal conductivity, 0.30±0.21 dB/cm-MHz for ultrasonic absorption coefficient and 1.63±0.46 dB/cm-MHz for ultrasonic attenuation coefficient.
Future studies should be focused on the optimum recipe of the atherosclerotic plaque phantoms that mimics the human atherosclerotic plaque (agar 4% w/v, gypsum 10% w/v and butter 10% w/v) and can be used for HIFU therapy. In future clinical trials the transducer (1-3mm wide) will be incorporated in a catheter which will be inserted intravascular, and then transferred to the heart arteries where ultrasonic ablation will take place for a minimum amount of time. This technology can be used in the future for clinical trials to treat plaques in the coronary arteries.
Subjects

HIFU

Ultrasound

Atherosclerosis

Cardiac treatment

Therapeutic Ultrasoun...

Atherosclerotic plaqu...

MRI

Attenuation

Absorption

Conductivity

Phantoms

File(s)
Thumbnail Image
Name

SOTIRIOU_PHD.pdf

Size

5.07 MB

Format

Adobe PDF

Checksum (MD5)

77f9bdb3c40172b338388f6f8dcbf3b9

Thumbnail Image
Name

ABSTRACT SOTIRIOU_PHD.pdf

Size

247.53 KB

Format

Adobe PDF

Checksum (MD5)

ab35b69e0fb50f40a6f7629b4f6a1fcd

Explore by
  • Collections
  • Research Outputs
  • Researchers
  • Faculty & Departments
  • Theses
  • Patents
  • Projects
  • Journals
  • Conferences
Useful Links
  • Researcher Portfolio Guide
  • Researcher Profile
  • Create an ORCID ID
  • CUT Open Access Author Fund
  • ETDS Guide
Copyright Policies

Use Sherpa/Romeo to find publisher copyright policies

Go
Go
  • SPARC Author Addendum Engine
  • National Open Access Policy in Cyprus
Deposit your work to Ktisis
  • Self-archiving. Please sign in to Ktisis.
  • Email your work to:
    library.dspace@cut.ac.cy
  • Contact your subject librarian

Member of

OpenAIREre3dataOpenDOARCOREDART
Cyprus University of Technology
Library and
Information
Services

Copyright © 2022 - Library and Information Services Feedback - Built with DSpace-CRIS - 4Science

  • Accessibility settings
  • Privacy policy
  • End User Agreement
COAR NotifyCOAR Notify