Τίτλος: Simple methods to test the accuracy of MRgFUS robotic systems
Συγγραφείς: Antoniou, Anastasia 
Drakos, Theocharis 
Giannakou, Marinos 
Evripidou, Nikolas 
Georgiou, Leonidas 
Christodoulou, Theodora 
Panayiotou, Natalie 
Ioannides, Cleanthis 
Zamboglou, Nikolaos 
Damianou, Christakis A. 
Major Field of Science: Engineering and Technology
Field Category: Electrical Engineering - Electronic Engineering - Information Engineering
Λέξεις-κλειδιά: MRgFUS robotic devices;Motion accuracy;Simple evaluation methods
Ημερομηνία Έκδοσης: Αυγ-2021
Πηγή: The International Journal of Medical Robotics and Computer Assisted Surgery, 2021, vol. 17, no. 4, articl. no. e2287
Volume: 17
Issue: 4
Περιοδικό: The International Journal of Medical Robotics and Computer Assisted Surgery 
Περίληψη: Background: Roboticassisted diagnostic and therapeutic modalities require ahighly accurate performance to be certified for clinical application. In this paper,three simple methods for assessing the accuracy of motion of magnetic resonanceguided focused ultrasound (MRgFUS) robotic systems are presented.Methods: The accuracy of motion of a 4 degrees of freedom robotic systemintended for preclinical use of MRgFUS was evaluated by calliperbased and mag-netic resonance imaging (MRI) methods, as well as visually by performing multipleablations on a plastic film.Results: The benchtop results confirmed a highly accurate motion in all axes ofoperation. The spatial positioning errors estimated by MRI evaluation were definedby the size of the imaging pixels. Lesions arrangement in discrete and overlappingpatterns confirmed satisfactory alignment of motion trajectories.Conclusions: We believe the methods presented here should serve as a standard forevaluating the accuracy of motion of MRgFUS robotic systems.
URI: https://hdl.handle.net/20.500.14279/23019
ISSN: 1478596X
DOI: 10.1002/rcs.2287
Rights: © The Authors. This is an open access article under the terms of the Creative Commons AttributionNonCommercial License.
Attribution-NonCommercial-NoDerivatives 4.0 International
Type: Article
Affiliation: Cyprus University of Technology 
Medsonic Ltd 
German Oncology Center 
Εμφανίζεται στις συλλογές:Άρθρα/Articles

Αρχεία σε αυτό το τεκμήριο:
Αρχείο Περιγραφή ΜέγεθοςΜορφότυπος
rcs.2287.pdfFulltext948 kBAdobe PDFΔείτε/ Ανοίξτε
CORE Recommender
Δείξε την πλήρη περιγραφή του τεκμηρίου

SCOPUSTM   
Citations

8
checked on 2 Φεβ 2024

WEB OF SCIENCETM
Citations

7
Last Week
0
Last month
0
checked on 29 Οκτ 2023

Page view(s)

247
Last Week
3
Last month
7
checked on 25 Ιουλ 2024

Download(s)

128
checked on 25 Ιουλ 2024

Google ScholarTM

Check

Altmetric


Αυτό το τεκμήριο προστατεύεται από άδεια Άδεια Creative Commons Creative Commons