Please use this identifier to cite or link to this item: https://hdl.handle.net/20.500.14279/1488
Title: Effect of a flow-streamlining implant at the distal anastomosis of a coronary artery bypass graft
Authors: Pedroso, Pedro D. 
Eleftheriou, Evangelos C. 
Anayiotos, Andreas 
metadata.dc.contributor.other: Αναγιωτός, Ανδρέας
Major Field of Science: Engineering and Technology
Field Category: ENGINEERING AND TECHNOLOGY
Keywords: Cameras;Computer simulation;Fetal heart rate monitoring;Arteriovenous anastomosis
Issue Date: Jul-2002
Source: Annals of Biomedical Engineering, 2002, vol. 30, no. 7, pp. 917-926
Volume: 30
Issue: 7
Start page: 917
End page: 926
Journal: Annals of Biomedical Engineering 
Abstract: Intimal thickening in the coronary artery bypass graft (CABG) distal anastomosis has been implicated as the, major cause of restenosis and long-term graft failure. Several studies point to the interplay between nonumiform hemodynamics including disturbed flows and recirculation zones, wall shear stress, and long particle residence time as possible etiologies. The hemodynamic features of two anatomic models of saphenous-vein CABGs were studied and compared. One simulated an anastomosis with both diameter and compliance mismatch and a curvature at the connection, analogous to the geometry observed in a conventional cardiothoracic procedure. The other, simulated an anastomosis with a flow stabilizing anastomotic implant connector which improves current cardiothoracic procedures by eliminating the distal vein bulging and curvature. Physiologic flow conditions were imposed on both models and qualitative analysis of the flow was performed with dye injection and a digital camera. Quantitative analysis was performed with laser Doppler velocimetry. Results showed that the presence of the bulge at the veno-arterial junction, contributed to the formation of accentuated secondary structures (helices), which progress into the flow divider and significantly affect radial velocity components at the host vessel up to four diameters downstream of the junction. The model with the implant, achieved more hemodynamically efficient conditions on the host vessel with higher mean and maximum axial velocities and lower radial velocities than the conventional model. The presence of the sinus may also affect the magnitude and shape of the shear stress at locations where intimal thickening occurs. Thus, the presence of the implant creates a more streamlined environment with more primary and less secondary flow components which may then inhibit the development of intimal thickening, restenosis, and ultimate failure of the saphenous vein graft.
URI: https://hdl.handle.net/20.500.14279/1488
ISSN: 00906964
DOI: 10.1114/1.1500407
Rights: © Springer
Attribution-NonCommercial-NoDerivs 3.0 United States
Type: Article
Affiliation: University of Alabama at Birmingham 
Affiliation : University of Alabama at Birmingham 
Appears in Collections:Άρθρα/Articles

CORE Recommender
Show full item record

SCOPUSTM   
Citations

18
checked on Nov 9, 2023

WEB OF SCIENCETM
Citations

15
Last Week
0
Last month
0
checked on Oct 20, 2023

Page view(s)

467
Last Week
2
Last month
27
checked on Apr 28, 2024

Google ScholarTM

Check

Altmetric


This item is licensed under a Creative Commons License Creative Commons