Τίτλος: Nanometer-sized ceria-coated silica-iron oxide for the reagentless microextraction/preconcentration of heavy metals in environmental and biological samples followed by slurry introduction to ICP-OES
Συγγραφείς: Dados, A 
Paparizou, E 
Eleftheriou, Polyvios 
Papastephanou, C 
Stalikas, C D 
Major Field of Science: Engineering and Technology
Field Category: Mechanical Engineering
Λέξεις-κλειδιά: Heavy metals microextraction/preconcentration; Inductively coupled plasma-optical emission spectrometry; Nanometer-sized magnetic ceria; Slurry suspension sampling;Inductively coupled plasma-optical emission spectrometry;Nanometer-sized magnetic ceria;Slurry suspension sampling
Ημερομηνία Έκδοσης: Απρ-2014
Πηγή: Talanta,2014, vol. 121, App. 127-135
Volume: 121
Start page: 127
End page: 135
Περιοδικό: Talanta 
Περίληψη: A slurry suspension sampling technique is developed and optimized for the rapid microextraction of heavy metals and analysis using nanometer-sized ceria-coated silica-iron oxide particles and inductively coupled plasma optical emission spectrometry (ICP-OES). Magnetic-silica material is synthesized by a co-precipitation and sol-gel method followed by ceria coating through a precipitation. The large particles are removed using a sedimentation-fractionation procedure and a magnetic homogeneous colloidal suspension of ceria-modified iron oxide-silica is produced for microextraction. The nanometer-sized particles are separated from the sample solution magnetically and analyzed with ICP-OES using a slurry suspension sampling approach. The ceria-modified iron oxide-silica does not contain any organic matter and this probably justifies the absence of matrix effect on plasma atomization capacity, when increased concentrations of slurries are aspirated. The As, Be, Mo, Cr, Cu, Pb, Hg, Sb, Se and V can be preconcentrated by the proposed method at pH 6.0 while Mn, Cd, Co and Ni require a pH ≥ 8.0. Satisfactory values are obtained for the relative standard deviations (2-6%), recoveries (88-102%), enrichment factors (14-19) and regression correlation coefficients as well as detectability, at sub-μg L(-1) levels. The applicability of magnetic ceria for the microextraction of metal ions in combination with the slurry introduction technique using ICP is substantiated by the analysis of environmental water and urine samples.
ISSN: 18733573
DOI: 10.1016/j.talanta.2013.12.045
Rights: © Elsevier
Type: Article
Affiliation: Cyprus University of Technology 
University of Ioannina 
Cp Foodlab Ltd. 
Εμφανίζεται στις συλλογές:Άρθρα/Articles

CORE Recommender
Δείξε την πλήρη περιγραφή του τεκμηρίου

SCOPUSTM   
Citations

36
checked on 1 Φεβ 2024

WEB OF SCIENCETM
Citations

35
Last Week
1
Last month
1
checked on 29 Οκτ 2023

Page view(s) 50

279
Last Week
3
Last month
12
checked on 10 Μαϊ 2024

Google ScholarTM

Check

Altmetric


Όλα τα τεκμήρια του δικτυακού τόπου προστατεύονται από πνευματικά δικαιώματα