Please use this identifier to cite or link to this item: https://hdl.handle.net/20.500.14279/13881
Title: A simulation model for predicting the temperature during the application of MR-guided focused ultrasound for stroke treatment using pulsed ultrasound
Authors: Damianou, Christakis A. 
Hadjisavvas, V. 
Major Field of Science: Engineering and Technology
Field Category: Electrical Engineering - Electronic Engineering - Information Engineering
Keywords: Duty factor;Focal depth;Pulsed ultrasound
Issue Date: 13-Oct-2011
Source: 10th International Symposium on Therapeutic Ultrasound, ISTU 2010; Tokyo; Japan; 9 June 2010 through 12 June 2010
Volume: 1359
Conference: AIP Conference Proceedings 
Abstract: In this paper a simulation model for predicting the temperature during the application of MR-guided focused ultrasound for stroke treatment using pulsed ultrasound is presented. A single element spherically focused transducer of 5 cm diameter, focusing at 10 cm and operating at either 0.5 MHz or 1 MHz was considered. The power field was estimated using the KZK model. The temperature was estimated using the bioheat equation. The goal was to extract the acoustic parameters (power, pulse duration, duty factor and pulse repetition frequency) that maintain a temperature increase of less than 1°C during the application of a pulse ultrasound protocol. It was found that the temperature change increases linearly with duty factor. The higher the power, the lower the duty factor needed to keep the temperature change to the safe limit of 1°C. The higher the frequency the lower the duty factor needed to keep the temperature change to the safe limit of 1°C. Finally, the deeper the target, the higher the duty factor needed to keep the temperature change to the safe limit of 1°C. The simulation model was tested in brain tissue during the application of pulse ultrasound and the measured temperature was in close agreement with the simulated temperature. This simulation model is considered to be very useful tool for providing acoustic parameters (frequency, power, duty factor, pulse repetition frequency) during the application of pulsed ultrasound at various depths in tissue so that a safe temperature is maintained during the treatment. This model could be tested soon during stroke clinical trials. © 2011 American Institute of Physics.
Description: AIP Conference Proceedings, Volume 1359
ISBN: 9780735409170
DOI: 10.1063/1.3607914
Rights: © AIP Publishing
Type: Conference Papers
Affiliation : Frederick University 
Medsonic Ltd 
City University London 
Publication Type: Peer Reviewed
Appears in Collections:Δημοσιεύσεις σε συνέδρια /Conference papers or poster or presentation

CORE Recommender
Show full item record

Page view(s) 10

275
Last Week
0
Last month
1
checked on Nov 21, 2024

Google ScholarTM

Check

Altmetric


Items in KTISIS are protected by copyright, with all rights reserved, unless otherwise indicated.