Title: Development and preliminary evaluation of a robotic device for mri-guided needle breast biopsy
Authors: Ioannides, Cleanthis 
Antoniou, Anastasia 
Zinonos, Vasiliki 
Damianou, Christakis A. 
Major Field of Science: Engineering and Technology
Field Category: Electrical Engineering - Electronic Engineering - Information Engineering
Keywords: accuracy;agar phantom;breast biopsy;MRI-guided;positioning device
Issue Date: 1-Jan-2024
Source: Journal of Medical Robotics Research, 2024
Journal: Journal of Medical Robotics Research 
Abstract: This study concerns the development and evaluation of a simple and ergonomic robotic system for Magnetic Resonance Imaging (MRI)-guided needle breast biopsy with lateral needle approach. The device comprises two piezoelectrically actuated linear motion stages intended to align a needle supporter with the target for manual needle insertion. The device demonstrated submillimeter accuracy and safe operation within a 3 T clinical MRI scanner. In phantom studies, tumor simulators of varying sizes were successfully targeted in both laboratory and MRI settings.