Παρακαλώ χρησιμοποιήστε αυτό το αναγνωριστικό για να παραπέμψετε ή να δημιουργήσετε σύνδεσμο προς αυτό το τεκμήριο: https://hdl.handle.net/20.500.14279/18054
Τίτλος: Performance characteristics of automated clinical chemistry analyzers using commercial assay reagents contributing to quality assurance and clinical decision in a hospital laboratory
Συγγραφείς: Tzortzopoulos, Athanasios 
Raftopoulos, Vasilios 
Talias, Michael A. 
Major Field of Science: Medical and Health Sciences
Field Category: Clinical Medicine
Λέξεις-κλειδιά: Clinical chemistry;Clinical decision;Evaluation;Sigma metrics;Uncertainty;Validation studies
Ημερομηνία Έκδοσης: 2-Ιαν-2020
Πηγή: Scandinavian Journal of Clinical and Laboratory Investigation, 2020, vol. 80, no. 1, pp. 46-54
Volume: 80
Issue: 1
Start page: 46
End page: 54
Περιοδικό: Scandinavian Journal of Clinical and Laboratory Investigation 
Περίληψη: Background: Clinical laboratories provide essential diagnostic services that are essential in clinical decision making, contributing to the quality of healthcare. The performance of two Siemens ADVIA 1800 analyzers was characterized in a hospital Biochemistry laboratory in order to evaluate the analytical characteristics of such automated analyzer systems using nonoriginal assay reagents attempting to support laboratory quality service and crucial clinical decision making. Methods: We independently completed performance validation studies including trueness, precision, sensitivity as well as measurement of uncertainty and sigma metrics calculation for 25 biochemical parameters. Results: Trueness expressed as bias was less than 20% for both ADVIA 1800 analyzers. Within run and total precisions expressed as CV% were ≤9.85% on both analyzers for most parameters studied with few exceptions (Mg, TB, DB, Cl, HDL and UA) observed either in low or in high level samples and between the two analyzers. LoB, LoD and LoQ values produced by the two analyzers were comparable except Cl. Uncertainty values produced by the two analyzers were comparable with no significant differences. Quality performance of reagent assays was studied using the sigma metrics system. The sigma values were plotted on normalized method decision charts for graphical representation of assay performances for each analyzer. Conclusions: The two ADVIA systems, independently evaluated, showed consistent performance characteristics with certain discrepancies by several reagents. Sigma analysis was helpful for revealing the quality performance of non-original reagents supporting the need for strict assessment of quality assurance and in some instances optimization/improvement of assay methods.
ISSN: 15027686
DOI: 10.1080/00365513.2019.1695282
Rights: © Medisinsk Fysiologisk Forenings Forlag (MFFF)
Type: Article
Affiliation: Cyprus University of Technology 
General Hospital of Agrinio 
Aghia Sophia Children's Hospital 
Open University Cyprus 
Εμφανίζεται στις συλλογές:Άρθρα/Articles

CORE Recommender
Δείξε την πλήρη περιγραφή του τεκμηρίου

SCOPUSTM   
Citations

2
checked on 1 Φεβ 2024

WEB OF SCIENCETM
Citations

1
Last Week
0
Last month
0
checked on 29 Οκτ 2023

Page view(s)

316
Last Week
3
Last month
7
checked on 1 Μαϊ 2024

Google ScholarTM

Check

Altmetric


Όλα τα τεκμήρια του δικτυακού τόπου προστατεύονται από πνευματικά δικαιώματα