Παρακαλώ χρησιμοποιήστε αυτό το αναγνωριστικό για να παραπέμψετε ή να δημιουργήσετε σύνδεσμο προς αυτό το τεκμήριο: https://hdl.handle.net/20.500.14279/1764
Τίτλος: Color Doppler velocity accuracy proximal to regurgitant orifices: influence of orifice aspect ratio
Συγγραφείς: Myers, Jerry G. 
Elmahdi, Abdelaziz M. 
Anayiotos, Andreas 
metadata.dc.contributor.other: Αναγιωτός, Ανδρέας
Λέξεις-κλειδιά: Computational fluid dynamics;Computational fluid dynamics;Computer simulation
Ημερομηνία Έκδοσης: 1999
Πηγή: Ultrasound in Medicine and Biology, 1999, Volume 25, Issue 5, Pages 771-792
Περίληψη: Many noninvasive methodologies used for the accurate evaluation of valvular regurgitation require precise velocity measurements from ultrasound instruments. Previous studies have indicated that velocity measurements from color Doppler (CD) instruments are susceptible to errors due to the interaction of the ultrasound beam and the proximal orifice flow field. This study examined the influence of high aspect ratio (AR) orifices on the CD velocity error. Center line velocity error distributions for orifices ranging from 7.07 to 78.5 mm2, varying in shape from circular to an AR = 8 ellipse, were evaluated using a numerical model of the ultrasound beam and the simulated regurgitant flow field. An in vitro study was also performed and confirmed the findings of the numerical model. The study showed that increasing AR does not significantly change the error characteristics. The study confirmed that orifice size is the dominant factor in the error distribution, and that corrections speculated for circular orifices can be extended to elliptical orifices without significant errors. Many noninvasive methodologies used for the accurate evaluation of valvular regurgitation require precise velocity measurements from ultrasound instruments. Previous studies have indicated that velocity measurements from color Doppler (CD) instruments are susceptible to errors due to the interaction of the ultrasound beam and the proximal orifice flow field. This study examined the influence of high aspect ratio (AR) orifices on the CD velocity error. Center line velocity error distributions for orifices ranging from 7.07 to 78.5 mm2, varying in shape from circular to an AR = 8 ellipse, were evaluated using a numerical model of the ultrasound beam and the simulated regurgitant flow field. An in vitro study was also performed and confirmed the findings of the numerical model. The study showed that increasing AR does not significantly change the error characteristics. The study confirmed that orifice size is the dominant factor in the error distribution, and that corrections speculated for circular orifices can be extended to elliptical orifices without significant errors.
URI: https://hdl.handle.net/20.500.14279/1764
ISSN: 03015629
DOI: http://dx.doi.org/10.1016/S0301-5629(99)00023-X
Rights: © 1999 World Federation for Ultrasound in Medicine & Biology
Type: Article
Affiliation: University of Alabama at Birmingham 
Εμφανίζεται στις συλλογές:Άρθρα/Articles

CORE Recommender
Δείξε την πλήρη περιγραφή του τεκμηρίου

SCOPUSTM   
Citations

4
checked on 13 Φεβ 2018

Page view(s)

542
Last Week
0
Last month
1
checked on 6 Νοε 2024

Google ScholarTM

Check

Altmetric


Όλα τα τεκμήρια του δικτυακού τόπου προστατεύονται από πνευματικά δικαιώματα