Please use this identifier to cite or link to this item: https://hdl.handle.net/20.500.14279/13870
Title: High intensity focused ultrasound ablation of kidney guided by MRI
Authors: Velev, O. 
Pavlou, M. 
Kyriakou, K. 
Trimikliniotis, M. 
Damianou, Christakis A. 
Major Field of Science: Engineering and Technology
Field Category: Electrical Engineering - Electronic Engineering - Information Engineering
Keywords: Cavitation;Cortex;Kidney;Lesion;MRI;Ultrasound
Issue Date: 1-Mar-2004
Source: Ultrasound in Medicine and Biology, 2004, vol. 30, no. 3, pp. 397-404
Volume: 30
Issue: 3
Journal: Ultrasound in Medicine and Biology 
Abstract: The effectiveness of magnetic resonance imaging (MRI) to monitor therapeutic protocols of high-intensity focused ultrasound (HIFU), in freshly excised pig kidney cortex is investigated. For high quality imaging, the pulse sequence fast spin echo (FSE) T1- and T2-weighted, and proton density were evaluated. For fast imaging, the pulse sequence T1-weighted fast spoiled gradient (FSPGR) was used. The main goal was to evaluate the MRI detection of large lesions (bigger than 1 cm × 1 cm × 1 cm) that is achieved by moving the transducer in a predetermined pattern. The contrast between lesion and kidney tissue is excellent with either T1-weighted or T2-weighted FSE. With T1-weighted FSE, the best contrast is observed for recovery time (TR) between 200 ms and 400 ms. With T2-weighted FSE best contrast can be achieved for echo time (TE) between 16 and 32 ms. T2-weighted FSE was proven as the best pulse sequence to detect cavitational activity. This advantage is attributed to the significant difference in signal intensity between air spaces and necrotic tissue. Air spaces appear brighter than thermal lesions. Therefore, for therapeutic protocols created using cavitational mode, T2-weighted FSE may be the optimum pulse sequence to use. The proton density pulse sequence does not provide any advantage over the T1- and T2-weighted pulse sequences. Using T1-weighted FSPGR, acquisition time as low as 5 s could be achieved. Good contrast and signal-to-noise ratio (SNR) are achieved with TR = 100 ms and flip angle between 75 to 90°. The above techniques were very successful in detecting large lesion volumes. © 2004 World Federation for Ultrasound in Medicine & Biology.
ISSN: 03015629
DOI: 10.1016/j.ultrasmedbio.2003.10.018
Rights: © Elsevier
Type: Article
Affiliation : Frederick Institute of Technology 
Ygia Polyclinic Hospital 
Intercollege 
Publication Type: Peer Reviewed
Appears in Collections:Άρθρα/Articles

CORE Recommender
Show full item record

SCOPUSTM   
Citations

34
checked on Feb 1, 2024

WEB OF SCIENCETM
Citations

42
Last Week
0
Last month
0
checked on Oct 29, 2023

Page view(s)

289
Last Week
0
Last month
4
checked on Dec 3, 2024

Google ScholarTM

Check

Altmetric


Items in KTISIS are protected by copyright, with all rights reserved, unless otherwise indicated.