Παρακαλώ χρησιμοποιήστε αυτό το αναγνωριστικό για να παραπέμψετε ή να δημιουργήσετε σύνδεσμο προς αυτό το τεκμήριο:
https://hdl.handle.net/20.500.14279/2447
Τίτλος: | Phase velocity mapping through a NiTi stent with different levels of stenosis | Συγγραφείς: | Holton, Andrea D. Venugopalan, Ramakrishna Anayiotos, Andreas |
metadata.dc.contributor.other: | Αναγιωτός, Ανδρέας | Λέξεις-κλειδιά: | Cardiovascular system;Flow meters;Magnetic resonance imaging | Ημερομηνία Έκδοσης: | 2003 | Πηγή: | ASME International Mechanical Engineering Congress, 2003, Washington, Olympia | Περίληψη: | Magnetic Resonance Imaging has the potential to become the new gold standard in the management of cardiovascular disease, particularly in the assessment of arterial lesions and post-intervention restenosis. NiTi stents exhibit superior MRI compatibility and allow direct MR imaging through the stented area. In the present study, pulsatile in-vitro flow measurements were obtained by Magnetic Resonance Phase-Velocity mapping (MR-PVW) through a nickel-titanium alloy stent with different grades of stenosis: a) 75%, 90% (symmetric) and 50% (asymmetric). The results showed clear visibility of the lumen through the stent/stenosis when compared with independent calculations of flow. The detection and evaluation of instent restenosis using these techniques could provide a non-invasive means of screening for further management of patients. | URI: | https://hdl.handle.net/20.500.14279/2447 | ISSN: | 10716947 | Rights: | © ASME | Type: | Conference Papers | Affiliation: | University of Alabama at Birmingham |
Εμφανίζεται στις συλλογές: | Δημοσιεύσεις σε συνέδρια /Conference papers or poster or presentation |
CORE Recommender
Όλα τα τεκμήρια του δικτυακού τόπου προστατεύονται από πνευματικά δικαιώματα