Τίτλος: Plastic scintillator-based fibre dosimeters for measurement of X-ray pulses in a clinical setting
Συγγραφείς: Kam, Wern 
Ioannou, Andreas 
Martyn, Michael 
Sullivan, F. J. 
Posporis, Andreas 
Woulfe, Peter 
Kalli, Kyriacos 
O’Keeffe, Sinead 
Major Field of Science: Engineering and Technology
Field Category: Electrical Engineering - Electronic Engineering - Information Engineering
Λέξεις-κλειδιά: Plastic scintillator;fibre dosimeters;X-ray pulses
Ημερομηνία Έκδοσης: 19-Μαΐ-2022
Volume: 12140
Περιοδικό: Proceedings of SPIE - The International Society for Optical Engineering 
Περίληψη: This work presents the development of plastic scintillator (BCF-10) based optical fibre sensors for medical radiotherapy dosimetry. Two different designs of BCF10 joined to PMMA (Polymethyl methacrylate) fibre were considered, based on simple Plug and Play designs for the rapid and effective assembly of radiation sensors. The first design was a simple butt-coupling arrangement sheathed in tubing, with an outer diameter of <2 mm. The second design explored the coupling joint of a cylindrical protrusion and hollow part of BCF10-PMMA that were achieved using femtosecond laser machining; the purpose of which was to maintain the original 1-mm fibre diameter for the sensor probe. The two fibres were pressed together and sealed with UV curing, hence the reference to a Plug and Play architecture. Both sensors exhibit higher output counts at the higher dose rate (due to the higher number of radiation pulses), although a discernible signal is observed at 50 MU/min for 6 MV, 15 MV energies and both sensors. When comparing both sensors with the different joint coupling designs, the flat surface connection of BCF-10 to PMMA demonstrates slightly higher photon counts compared with the micro-machined sensor (Plug n Play). However, the difference is small and the Plug n Play sensor benefits from the smaller sensor diameter (1 mm diameter), which is suitable for inserting into a small applicator or in-vivo monitoring. In the second section, micro-pulses of X-Ray radiation from Siemens Linear Accelerator (linac) were obtained and compared for two different energies and dose rates. Both of the sensors demonstrate the feasibility to be used for characterisation of X-ray pulses from a clinical linac.
URI: https://hdl.handle.net/20.500.14279/29543
ISBN: 9781510651562
ISSN: 0277786X
DOI: 10.1117/12.2624516
Rights: © Society of Photo-Optical Instrumentation Engineers (SPIE)
Attribution-NonCommercial-NoDerivatives 4.0 International
Type: Conference Papers
Affiliation: University of Limerick 
Galway Clinic 
Cyprus University of Technology 
Εμφανίζεται στις συλλογές:Δημοσιεύσεις σε συνέδρια /Conference papers or poster or presentation

CORE Recommender
Δείξε την πλήρη περιγραφή του τεκμηρίου

SCOPUSTM   
Citations 50

1
checked on 14 Μαρ 2024

Page view(s)

96
Last Week
3
Last month
8
checked on 11 Μαϊ 2024

Google ScholarTM

Check

Altmetric


Αυτό το τεκμήριο προστατεύεται από άδεια Άδεια Creative Commons Creative Commons