Please use this identifier to cite or link to this item: http://ktisis.cut.ac.cy/handle/10488/9908
Title: In situ disdrometer calibration using multiple DSD moments
Authors: Lane, John E. 
Kasparis, Takis 
Metzger, Philip T. 
Jones, W. Linwood 
Keywords: DSD moments
Impact disdrometer
In situ calibration
Optical extinction
Rainfall rate
Issue Date: 1-Jan-2014
Publisher: Versita
Source: Acta Geophysica Volume 62, Issue 6, 2014, Pages 1450-1477
Abstract: In situ calibration is a proposed strategy for continuous as well as initial calibration of an impact disdrometer. In previous work, a collocated tipping bucket had been utilized to provide a rainfall rate based ∼11/3 moment reference to an impact disdrometer’s signal processing system for implementation of adaptive calibration. Using rainfall rate only, transformation of impulse amplitude to a drop volume based on a simple power law was used to define an error surface in the model’s parameter space. By incorporating optical extinction second moment measurements with rainfall rate data, an improved in situ disdrometer calibration algorithm results due to utilization of multiple (two or more) independent moments of the drop size distribution in the error function definition. The resulting improvement in calibration performance can be quantified by detailed examination of the parameter space error surface using simulation as well as real data.
URI: http://ktisis.cut.ac.cy/handle/10488/9908
ISSN: 18956572
Rights: © 2014, Versita Warsaw and Springer-Verlag Wien.
Appears in Collections:Άρθρα/Articles

Files in This Item:
File Description SizeFormat 
Kasparis T..pdf1.46 MBAdobe PDFView/Open
Show full item record

Page view(s) 50

27
Last Week
0
Last month
3
checked on Aug 23, 2017

Download(s) 50

7
checked on Aug 23, 2017

Google ScholarTM

Check


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.