Please use this identifier to cite or link to this item: http://ktisis.cut.ac.cy/handle/10488/7505
Title: The role of vascular calcification in inducing fatigue and fracture of coronary stents
Authors: Halwani, Dina O. 
Anderson, Peter G. 
Anayiotos, Andreas 
Keywords: Calcification;Biomechanics;Biomineralization
Category: Medical Engineering
Field: Engineering and Technology
Issue Date: 2012
Publisher: John Wiley & Sons
Source: Journal of Biomedical Materials Research - Part B Applied Biomaterials, 2012, Volume 100 B, Issue 1, Pages 292-304
Abstract: Traditional approaches for in-vitro pulsatile and fatigue testing of endovascular stents do not take into consideration the pathologies of the stented vessel and their associated biomechanical effects. One important pathology is calcification, which may be capable of inducing changes in the vessel wall leading to inhomogeneous distribution of stresses combined with wall motion during the cardiac cycle. These local property changes in the region adjacent to stents could directly influence in-vivo stent performance. Seven cases containing a total of 18 stents were obtained from autopsy. Radiographs were evaluated and vessels were sectioned for histology and stent topographical analysis. Stents were retrieved by chemical removal of surrounding tissue and surfaces were evaluated using 3D digital optical and scanning electron microscopy for biomechanical abrasion and fracture features. Pathologic complications such as restenosis and thrombus formation were assessed from histological sections. Direct evidence of fracture was found in 6 of the 7 cases (in 12 out of 18 stents; 9 drug eluting and 3 bare metal). The degree of stent alterations was variable, where separation of segments due to fracture occurred mostly in drug-eluting stents. All fracture surfaces were representative of a high cycle fatigue mechanism. These fractures occurred in complex lesions involving the presence of diffuse calcification alone, or in combination with vessel angulations and multiple overlapping stents. Morphologic analysis of tissue at or near some fracture sites showed evidence of thrombus formation and/or neointimal tissue growth.
URI: http://ktisis.cut.ac.cy/handle/10488/7505
ISSN: 15524973
DOI: 10.1002/jbm.b.31911
Rights: © 2011 Wiley Periodicals, Inc
Type: Article
Appears in Collections:Άρθρα/Articles

Show full item record

SCOPUSTM   
Citations

21
checked on Dec 13, 2018

WEB OF SCIENCETM
Citations 5

20
checked on Feb 15, 2019

Page view(s)

46
Last Week
1
Last month
2
checked on Feb 20, 2019

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.