Title: Acoustic and thermal characterization of agar based phantoms used for evaluating focused ultrasound exposures
Authors: Menikou, Georgios 
Damianou, Christakis A. 
Keywords: Ultrasound;Agar;Attenuation;Conductivity
Category: Other Medical Sciences
Field: Medical and Health Sciences
Issue Date: 1-Jun-2017
Publisher: BIOMED CENTRAL LTD
Source: JOURNAL OF THERAPEUTIC ULTRASOUND, Volume: 5, Article Number: 14, Published: JUN 1 2017
metadata.dc.doi: http://dx.doi.org/10.1186/s40349-017-0093-z
Journal: Journal of Therapeutic Ultrasound
Abstract: Background This study describes a series of experimental work completed towards characterizing candidate materials for fabricating brain and muscle tissue mimicking phantoms. Methods The acoustic speed, attenuation, impedance, thermal diffusivity, specific heat and thermal conductivity were measured. Results The resulting brain (2% w/v agar-1.2% w/v Silica Dioxide-25%v/v evaporated milk) and muscle tissue recipe (2% w/v agar-2% w/v Silica Dioxide-40%v/v evaporated milk) introduced a total attenuation coefficient of 0.59 dB/cm-MHz and 0.99 dB/cm-MHz respectively. Acrylonitrile Butadiene Styrene (ABS) possessed an attenuation coefficient of 16 dB/cm at 1 MHz which was found within the very wide range of attenuation coefficient values of human bones in literature. The thermal conductivity of the brain tissue phantom was estimated at 0.52 W/m°C and at 0.57 W/m.°Cfor the muscle. These values demonstrated that the proposed recipes conducted heat similar to the majority of most soft tissues found from bibliography. The soft tissue phantoms were also evaluated for their thermal repeatability after treating them repeatedly at different locations with the same sonication protocol and configuration. The average coefficient of variation of the maximum temperature at focus between the different locations was 2.6% for the brain phantom and 2.8% for the muscle phantom. Conclusions The proposed phantom closely matched the acoustic and thermal properties of tissues. Experiments using MR thermometry demonstrated the usefulness of this phantom to evaluate ultrasonic exposures.
URI: http://ktisis.cut.ac.cy/handle/10488/10543
ISSN: 2050-5736
Rights: © The Author(s). 2017
Type: Article
Appears in Collections:Άρθρα/Articles

Files in This Item:
File Description SizeFormat 
document.pdf2.77 MBAdobe PDFView/Open
Show full item record

Page view(s)

12
Last Week
2
Last month
checked on Dec 13, 2017

Download(s) 50

2
checked on Dec 13, 2017

Google ScholarTM

Check


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.